Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
5-1
5
5
Monitoring SpO
2
Pulse oximetry is a noninvasive method of continuously measuring oxygen 
saturation (SpO
2
) in arterial blood. The resultant SpO
2
 reading indicates the 
percentage of hemoglobin molecules in the arterial blood which are saturated 
with oxygen. SpO
monitoring is one of the tools available to assist in assess-
ing a patient’s cardiac and respiratory systems. This chapter explains how 
pulse oximetry works and describes how to use the HeartStart XLT to monitor 
SpO
2.
SpO
monitoring is always available in Manual Mode (if the option is pur-
chased). In AED Mode, SpO
monitoring is only available if SpO
2
 is config-
ured to 
2Q. 
For information on printing, storing, and retrieving patient information 
acquired while monitoring, see Chapter 9. 
WARNING
Do not rely solely on SpO
2
 readings; assess the patient at all times. SpO
2
 readings 
may be inaccurate in the presence of significant levels of carboxyhemoglobin or 
methemoglobin, in patients with restricted blood flow to the extremities (such as 
those in severe shock or hypothermia), or in the presence of excessive motion.