Selecting a Sensor
5-4
Monitoring SpO
2
The most important factor when selecting a sensor is the position of the light 
emitting diodes in relation to the photodetector; when a sensor is applied, the 
diodes and the photodetector must be opposite each other. Sensors are 
designed for patients within a specific weight range and for specific sites. Be 
sure to:
z
Select a sensor appropriate for the patient’s weight.
z
Select a sensor site with adequate perfusion.
z
Avoid application to sites with edematous tissue.
Reusable Sensors
Reusable sensors may be reused on different patients after they have been 
cleaned and disinfected (see the manufacturer’s instructions supplied with the 
sensor). 
Disposable Sensors
Disposable sensors should be used only once and then discarded. They can be 
relocated to a different application site on the patient if the first location does 
not give the desired results. Disposable sensors must not be reused on differ-
ent patients. 
Semi-disposable Sensors
Semi-disposable sensors can be reused, but the adhesive wrap must be dis-
carded after each use. Semi-disposable sensors are recommended for single-
patient use only.