Setting Alarms
5-8
Monitoring SpO
2
Setting Alarms
An alarm may be set to alert you if the SpO
2
 value falls below a specified 
lower limit. Lower limit alarm choices are 
 (no alarm), 90, 85, or 80. Press 
 repeatedly to cycle through the choices. Stop when the desired choice is 
displayed. A 
 appears in three seconds, indicating that the selected alarm is 
active. To review the alarm limit, press 
.
WARNING
SpO
2
 alarms are temporarily suspended in AED Mode during ECG analysis or when 
 is pressed (for the duration of the paused period). SpO
2
 alarms are also sus-
pended while charging for defibrillation and delivering a shock. 
Responding to an Alarm
When the SpO
value falls below the alarm limit, a continuous tone alerts you 
and the SpO
2
 value is displayed in inverse video.
Figure 5-4  SpO
2
 Alarm Triggered 
Press 
 to turn off the alarm. Refer to “Setting Alarms” if subsequent 
alarms are desired.