Monitoring
5-7
5
Monitoring
To monitor SpO
2
:
1. If the HeartStart XLT is not on, press On.
2. Apply the appropriate sensor to the patient.
3. Make sure the sensor cable is connected to the HeartStart XLT.
4. Press 
 to turn on SpO
2
 monitoring. 
A dashed line (---) is displayed under 
, while the oxygen saturation is 
measured and an SpO
2
 value is calculated. In a few seconds the SpO
2
 value is 
displayed in place of the dashed line. As the patient’s oxygen saturation 
changes, the SpO
2
 value is updated continuously.
Figure 5-3  SpO
2
 Monitoring Display 
To the right of the SpO
2
 value, a pleth bar and SpO
2
 alarm indicator are dis-
played. The pleth bar should be observed for fluctuation. It is an indication of 
pulsation detected by the sensor. The pleth bar should not be used as the sole 
indicator of pulsation because it can be influenced by movement and artifact. 
The 
 symbol indicates no alarm is set. 
Below the SpO
2
 value is the pulse rate derived from the pulse oximetry. 
6S2

6S2

PAUSE
HR ALARM
LEAD
SELECT
ANALYZE
Shocks: 3
00:00:49
Monitoring Rhythm
78
Pads
SpO2
98
PR 62