INSTALLATION AND SETUP 
2-29
MON2000
JULY 2010
 Set Up GC Directory
GC Controller system interface board) for 
the GC unit.
To successfully connect to the GC unit, the 
COM IDs specified via the File>GC 
Directory and the Application>Serial Ports 
dialogs must match. 
more information.
7. Click the Comments cell to enter any helpful 
information regarding a particular GC unit, 
such as location or purpose.
8. Use the dynamic pull-down menus to select 
the desired PC PortBaud RateData Bits
Stop BitsParity, and Handshaking 
settings.
9. Use the provided cells to input RTS On 
DelayRTS Off Delay, and Extra Delay 
values.
The Extra Delay field accepts numerical 
values (0 to 9000 milliseconds) for 
additional time to be added to the current 
communication timeout delay.
10.Use the Protocol pull-down menu to select 
the desired Modbus communications 
protocol.
11.If you plan to connect remotely to the GC 
unit via a modem or Ethernet connection, 
use the Connection Type (Direct/Remote) 
The following parameters support an auto-detect 
setting: Baud Rate, Data Bits, Stop Bits, Parity, 
Protocol.
Set any of these parameters to “?” (autodetect) and 
MON2000 will conform to the settings used by the GC 
unit.