Pressure Systems, Inc.                                                                                        Model 9046 User’s Manual 
 
 
 
Page 48 
 
www.PressureSystems.com 
Note
 
NUSS automatically creates a default Run State 0 for every module 
that uses its Run function. This run state uses the above sub-
command to configure three (3) "unbounded" autonomous data 
streams that the module will deliver concurrently at various rates to 
the host:  
•  Stream 1 contains Primary Transducer E.U. data 
(Temperature/Resistance) only for every channel and 
repeats every 1/2 second.   
•  Stream 2 contain a Thermal Alarm Prefix and Primary 
Transducer Counts and Volts data for every channel, and 
repeats every 2 seconds.   
•  Stream 3 contains Other (UTR/Source Voltage) Transducer 
data (in E. U., Counts, and Volts) for every channel and 
repeats every 7.5 seconds.  
The Run function will acquire all the data defined in the Run State, 
but will display only the data in the arriving streams that match the 
form’s currently configured Display requirements (called its Display 
Set).  If a user requires other stream contents or rates, these can 
be configured (as Run States 1-9) using the NUSS Run State 
Editor.  Any run-state can be assigned to the Run form of each 
module.