Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-42 
Where: 
name 
= name of the variable 
value 
= current value of variable 
warning_lo 
= lower limit warning (displayed if applicable) 
warning_hi 
= upper limit warning (displayed if applicable) 
data_lo  
= lower limit of allowable values 
data_hi  
= upper limit of allowable values 
 
Variables can be changed. Before changing the settings on any variables, please make sure you 
understand the consequences of the change. We recommend you call the factory before changing 
the settings on any variables. The general format for changing the settings on a variable is: 
V LIST name[=value [warn_lo [warn_hi]]] 
For example to change the warning limits on the box temperature type: 
V BOX_SET 30 10 50 
and the CPU should respond with: 
V DDD:HH:MM IIII BOX_SET=30 10 50(0-60) 
The CONFIG command lists the software configuration. 
For example:  
V CONFIG
 
 list software configuration 
The format of this listing is shown in the example below. 
CONFIG[0]
 = Revision H.7 
CONFIG[1]
 = NO
x
 Analyzer 
CONFIG[2]
 = SBC40 CPU