Teledyne 200AH Manuale Utente

Pagina di 196
Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
6-2 
6.2. 
  Zero/Span Valves Option 
The Zero/Span Valve option consists of a manifold of stainless steel valves plus drive electronics 
connected to the CPU. The Option consists of two variations. Option 50 has 2 - 2-way valves, 
1 3-way valve and operates in 3 modes: 
1.  Sample mode, all 3 valves off, sample gas flows through the sample port to the instrument. 
2.  Zero mode, sample-cal valve is energized to shut off sample flow, zero valve is energized to 
input zero gas. 
3.  Span mode, sample-cal valve is energized to shut off sample flow, span valve is energized to 
input low concentration range span gas. 
Option 52 consists of all of the features in Option 50 plus a second span valve to allow a low and 
high concentration of span gas. 
See  
Note:  The Zero/Span Valves option cannot be included in an instrument that also includes an Oxygen 
Sensor. 
Table 6-1:  Zero/Span Valve Operation 
Mode 
Description 
Reference Section 
1. 
Front panel operation via 
CALS and CALZ buttons 
2. 
Automatic operation using 
AUTOCAL 
3. 
Remote operation using the 
RS-232 interface 
4. 
Remote operation using 
external contact closures 
Section 7.5 - Automatic operation using external contact 
closures. Table 7-12 and Section 9.3.4.3 
 
Zero and span gas inlets should supply their respective gases in excess of the demand of the 
Analyzer. Supply and vent lines should be of sufficient length and diameter to prevent back 
diffusion and pressure effects. See Figure 2-4 for fitting location and tubing recommendations. 
Note that Zero/Span Valves cannot be included in an instrument that also includes the Oxygen 
sensor option.