Diagnostic Mode
5-8 
Configuring the Power-up Defaults Menu  
System A/D 
Values
The System A/D screen displays the current value of each analog-to-digital (A/D) channel in 
volts. Some of the channels are for AC-coupled signals (such as ECG input), so the numbers 
on the screen are constantly changing when an input signal is present. These AC-coupled 
values are shown to give an indication as to whether basic functionality of the channel is 
present, but no significance can be derived from the values of the numbers displayed.
The Primary and Secondary Status messages from the SpO
module are displayed and 
updated at the rate of about once per second. Presence of the correct SpO
message indicates 
that, at a basic level, communication between the SpO
module and the main monitor 
processor is working correctly. None of the displayed values can be changed or reset in this 
screen. When in the System A/D screen, the Return option is always highlighted. Press the 
wheel to return to the Diagnostic Menu. Rotating the wheel while in the System A/D screen 
has no effect. The A/D channel designators are shown in the following table.
      
           
DIAGNOSTIC MENU
  
                                       
    
SYSTEM A/D VALUES
1.                                                         12.
2.
                                                            13.
3.                                                         14.
4.                                                         15.
5.                                                         16.
6.                                                         17.
7.                                                         18.
8.                                                         19.
9.                                                         20.
10.                                                       21.
11.
             
Return
SpO2 S1         SpO2 Alive           SpO2 S2        No Sensor