Diagnostic Mode
5-10 
Configuring the Power-up Defaults Menu  
NBP Test
A NBP Test screen is provided to facilitate troubleshooting problems and performing 
verification testing for the NBP subsystem. Typically, when these tests are performed, the 
pneumatic system is connected to an external pressure-reading device and a closed reference 
volume. The NBP Test screen provides a real-time numeric display of the pressure in the 
pneumatic system, means for controlling the pump and valves are open or closed.
Warning
Warning
A blood pressure cuff, connected to the monitor, should never be applied to a human 
being while the monitor is in Diagnostic Mode. Injury could result.
The NBP Test screen elements are described below:
Menu Item
Function
Pressure (mmHg)
Displays the real-time value of the 
system pneumatic pressure in mmHg. 
The value is updated at the rate of 
approximately two times per second.
Proportional Valve
Indicates whether the proportional value 
is open or closed.
Safety Valve
Indicates whether the safety valve is 
open or closed.
      
           
DIAGNOSTIC MENU
  
                                       
             
NBP TEST
                                                   
 
Press NBP to activate pump;
release to stop pump.
Press Volume to open both valves;
release to close valve.
Press Alarm Silence to open proportional valve
and deflate at 3mmHg/s;
Pressure (mmHg) 0                    Proportional Valve: Open
                                             
Proportional Valve: Open
 
Return
release to close valve
Press Contrast to perform