Diagnostic Mode
5-12 
Configuring the Power-up Defaults Menu  
CO
2
 Test
The CO
2
 Test Screen is provided to facilitate troubleshooting problems and performing 
verification testing to ensure the measuring results are accurate. See “CO2 Performance” on 
page 4-29
 for i
nformation on how to perform these tests. 
Warning
Warning
A FilterLine, connected to the monitor, should never be applied to a human being while 
the monitor is in Diagnostic Mode. Injury could result.
      
           
DIAGNOSTIC MENU
  
                                       
             
CO2 TEST
 
                                       
CO2 Pressure in mmHg (ambient/cell):     745/740
Pump Voltage (Volts):                                3.0
CO2 (mmHg):                                            0.00
Status:
Set Barometric Pressure (mmHg):             745
Flow Rate Check and Calibration:
CO2 Calibration:                                07/16/2002 15:13
Pump Op Time:                                           58