Philips 862474 ユーザーズマニュアル

ページ / 182
Measuring and Displaying Parameters
Specifications  
12-7
NBP (Non-
Invasive 
Blood 
Pressure)
Temperature
Parameter 
Specification
Standards
Meets performance standards of ANSI/
AAMI SP10-1992
Technique
Oscillometric
Cuff Pressure Display
10 to 300 mmHg
Pulse Rate Range
40 to 200 b/min
Blood Pressure Accuracy
Mean error and standard deviation per 
ANSI/AAMI SP10-1992
a
a. Systolic and Diastolic blood pressure measurements determined with this device are 
equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation 
method, and mean pressure measurements determined with this device are equivalent to 
those obtained by an intra-arterial blood pressure measurement device, within the limits 
prescribed by the American National Standard, Electronic or automated 
sphygmomanometers.
Pulse Rate Accuracy
Greater of ± 2 BMP or ± 2%
Initial Cuff Inflation
180 mmHg
Subsequence Cuff Inflation
Prev SYS ±30 mmHg
Blood Pressure Measurement Range
Systolic
60 to 250 mmHg
Diastolic
40 to 220 mmHg
Mean Arterial Pressure
45 to 235 mmHg
Measurement Modes
Auto
Automatic measurements at intervals of 
1, 3, 5, 10, 30, 60 and 90 minutes
Manual
Single measurement initiated by Start/
Stop button
STAT
Series of consecutive measurements for 
5 minutes
Parameter 
Specification
Technique
Probe
Range
15 to 45
o
C (59 - 113
o
F)
Accuracy
± 0.1
o
C