Measuring and Displaying Parameters
Specifications  
12-9
CO
2
Complies with EN 864/ISO 9918
 
Parameter 
Specification
Range
0 to 98 mmHg (0 to 13 kPa), or 13% 
CO
2
 whichever is lower.
Flow Rate
50 ±7.5 ml/min
Warm-up Time
180 seconds max.; typically 30 seconds
Gas Sampling Delay Time
2.7 seconds
Resolution
Numeric
1.0 mmHg (0.1 kPa)
Wave
0.1 mmHg (0.01 kPa)
Rise Time
190 ms for adult mode (measured with 
sample line for humidified ventilation 
and airway adapter for adult)
Calibration Interval
4,000 operating hours
Auto Zero Interval
< 2% duty cycle. Increases if the 
temperature change by 8
o
C from the 
previous Auto-Zero interval, or the 
pressure changes by 45 mmHg.
Leak Tightness
< 40 mbar/min with 30% vacuum
Accuracy
a
a. Accuracy applies for respiration rate up to 80 b/min. For respiration rate above 80 b/min, the 
accuracy complies with EN 864/ISO 9918 (4 mmHg or ±12% of indicated value, whichever 
is greater) for End-tidal CO
2
 values exceeding 19 mmHg. The above accuracy is maintained 
to within 4% for the following gas mixtures: CO
2
: 0 - 13%; N
2
: 0 - 97.5%; O
2
: 0 - 100%; 
N
2
O: 0 - 80%; H
2
O: Dry to saturated. 
0 - 20 minutes
0 to 38 mmHg ±4 mmHg; 38 to 99 
mmHg ±12% of indicated value
> 20 minutes
0 to 38 mmHg ±2 mmHg 
38 to 99 mmHg ±5% of indicated value 
+ 0.08% for every 1.0 mmHg above 40 
mmHg.