Philips 862474 ユーザーズマニュアル

ページ / 182
Test Map
Testing the Monitor  
4-7
Test Map
The Test Map shows which tests are required in which situations. 
Service Event
(When performing...)
Test Block Required
(...complete these test)
Installation
• Visual 
• Power On 
Monitor exchange
• Visual 
• Power  On
Monitor opened
• Power On
• Pneumatic Leakage Test
• All Safety tests
Battery replacement
• Power On
• Battery Performance
• All Safety tests
Speaker replacement
• Power On
• Pneumatic Leakage Test
• All Safety tests
Backlight tube replacement
• Power On
• Pneumatic Leakage Test
• All Safety tests
NBP pump replacement
• Power On
• Pneumatic Leakage Test
• All Safety tests
• Preventative Maintenance Tests
SpO
2
 module replacement
• Power On
• Pneumatic Leakage Test
• All Safety tests
• SpO
2
 Tests
CO
2
 module replacement
• Power On
• Pneumatic Leakage Test
• All Safety tests
• CO
2
 Tests
Power supply module replacement
• Power On
• Pneumatic Leakage Test
• Safety Tests
Patient monitoring I/O module 
replacement
• Power  On
• Pneumatic Leakage Test
• All Safety tests
• All Performance tests 
Communications module replacement
• Power On
• Pneumatic Leakage Test