Philips 862474 ユーザーズマニュアル

ページ / 182
Equipment
4-8 
Testing the Monitor  
Equipment
The following table lists the equipment required for performance verification.
Equipment
Description
Digital multimeter (DMM)
Fluke model 87 or equivalent
Defib Sync Cable and
Switchcraft 3.5 mm phone plug
M4820A (see page 4-38 to use a substitute cable)
Switchcraft 750 or equivalent
SpO
2
 adapter cable
M1943A
SpO
reusable sensor, adult finger 
M1191A
ECG trunk cable
M1540C (ICU, AAMI) or M1550C (ICU, IEC)
ECG electrodes
Standard
ECG lead sets
M1605A (AAMI) or M1615A (IEC)
NBP tubing
M1599B
NBP reusable cuff, adult
40401C
Pulse oximeter tester
Clinical Dynamics Corp - SmartSat simulator with 
Nellcor simulator cable
ECG simulator
Dynatech Nevada medSim 300 or equivalent
NBP simulator
Bio-Tek BP Pump 2 or equivalent
Temperature simulator
MedSim 300 or equivalent
Respiration simulator
MedSim 300 or equivalent
Safety analyzer
Bio-Tek 601 Pro or equivalent
Stopwatch
Manual or electronic
Screwdriver
flathead and philips
Tweezers
any model
Electronic flowmeter
M1026-60144
Cal 1 gas (5% CO
2
)
15210-64010
Cal 2 gas (10% CO
2
)
15210-64020
Cal gas flow regulator
M2267A
Cal tube
13907A