Patient Safety Tests
4-42 
Testing the Monitor  
Enclosure
Leakage
Current
This test is compliance with EN60601-1 (Enclosure Leakage Current). Test at 110% of the 
nominal line voltage.
The analyzer leakage current indication must not exceed the values listed below.
Table 4-5.  Enclosure Leakage Current
EN60601-1 (1990 + A1, A2, A11, A12, A13) and UL2601 (2
nd
 Ed. 1997) US Deviations
What to record on your service record:
S(3): P or S(3): F
Step
Action
1
Connect the AC mains power cord to the analyzer as recommended by the 
analyzer operating instructions.
2
Using the appropriate test cable, connect the analyzer to either of the screws on 
the back of the monitor, next to the handle.
3
Turn the monitor on.
4
Perform the test as recommended by the analyzer operating instructions.
Test Condition
Polarity
Allowable Leakage Current
Normal
Normal
100 µA
Reversed
100 µA
S.F.C
1
 Open
Protective Earth
Normal
300 µA
Reversed
300 µA