Patient Safety Tests
Testing the Monitor  
4-43
Patient
Leakage
Current
This test measures patient leakage current in accordance with EN60601-1, Clause 19, for 
Class I, type CF equipment. Patient leakage current in this test is measured from any 
individual patient connection to earth (power ground).
This test requires a sample patient cable for each device parameter. These must be configured 
as recommended by the safety analyzer operating instructions.
Table 4-6.  Patient Leakage Current Values
What to record on the service record:
S(4): P or S(4): F
Step
Action
1
Configure the electrical safety analyzer as recommended by the analyzer 
operating instructions.
2
Connect the monitor’s AC mains power cord to the analyzer as recommended by 
analyzer operating instructions.
3
Connect the ECG test cable between the ECG connector on the monitor and the 
appropriate input connector on the analyzer.
4
Turn the monitor On.
5
Perform the test as recommended by the analyzer operating instructions.
Patient leakage current is measured under various conditions of the AC mains 
and protective earth conductor. For each condition, the measured leakage current 
must not exceed that indicated below.
6
Repeat the test for SpO
2
 and temperature patient connections, using 
appropriate test cables.
Test Condition
Polarity
Allowable Leakage 
Current (Max.)
Type CF
Normal
Normal
10 µA
Reversed
10 µA
S.F.C.(Open Earth/Ground)
Normal
50 µA
Reversed
50 µA