Teledyne AX300-I ユーザーズマニュアル

ページ / 47
Portable Oxygen Analyzer  
Operation 
 
 
 Teledyne 
Analytical 
Instruments 
17 
2.1.2.1
 
V-M
OUNT 
A
DAPTER 
I
NSTALLATION
 
The V-Mount Adapter consists of a matching plastic plate with 
integral V-grooves that attach to the rear of the instrument.   
To install V-Mount Adapter remove battery compartment door by 
prying up the hinged latch at the bottom of the cover, then slide the 
adapter plate into grooves provided in rear case. Replace the battery 
compartment door and secure door latch. See Figure 2-4. 
Note:  The door latch is a tight fit onto the battery cover. Use a 
coin to gently pry up the latch. 
Figure 2-4: V-Mount Adapter Installation 
2.1.2.2
 
U
NIVERSAL 
M
OUNTING 
C
LAMP 
I
NSTALLATION
 
The Universal Mounting Clamp is supplied with a 1/4-20 screw for 
securing the clamp to the rear of the instrument. A threaded brass insert is 
installed on the back of the instrument for this purpose. See Figure 2-5. 
Figure 2-5: Brass Insert for Universal Mounting Clamp 
2.1.3 Battery Installation 
Note:  Three “AA” alkaline batteries must be installed in the unit 
before the analyzer will operate. The unit must be 
recalibrated whenever new batteries are installed
.  
To install the batteries: