A&D AD-4405 ユーザーズマニュアル

ページ / 80
AD-4405 Weighing Indicator
 Page 
21 
 
 
5.6.
5.6.
5.6.
5.6.     Calibration Error Cod
Calibration Error Cod
Calibration Error Cod
Calibration Error Code List
e List
e List
e List    
 
Exiting from a Calibration error 
 
ESC
 key..... The key to return to the point where an error occurred. Retry the operation. 
 
 
ESC
 key while pressing the 
ON/OFF
 key.   
 
 
No parameters are changed, CanCel is displayed and the calibration mode 
is finished. Press the 
ON/OFF
 key to turn the display off. 
 
Error Code List 
 
  If an error has occurred during the calibration mode, the following code is displayed. 
Error code 
Description 
err 0 
In multi-interval scale. The last division is set to maximum ( d-50 ). 
Therefore the next division can not be entered. 
err 1 
Resolution exceeds 40000. (Resolution = maximum capacity/ minimum 
division) 
Reduce the maximum capacity or increase the minimum division. 
err 2 
err 3 
Load cell output is too large or too small at zero calibration. Check the weighing 
unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input 
Sensitivity". 
err 4 
Measuring calibration mass, the value exceeded maximum capacity. Reduce 
the calibration mass. 
err 5 
The selected calibration mass is smaller than the minimum division. 
err 6 
The new input sensitivity is less than 0.2 µV/division. Increase the input 
sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity". 
err 7 
Placing a mass on the weighing unit, the load cell output becomes a negative 
value. Check the load cell cable connections and the direction of load cell 
mounting. 
err 8 
The load cell output exceeds the input range before the maximum capacity. 
Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input 
Sensitivity". Replace with a load cell designed for a smaller output. Reduce 
maximum capacity. 
err 9 
The weight value is out of the input range at zero calibration or span 
calibration. Check the weighing unit and cables. 
err 12  The first weighing range is larger than second weighing range. 
err 13 
An incorrect mass was selected at the digital linearization function. 
Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3. 
g err 
An unacceptable value was selected in the gravity acceleration function. 
 
 
 
 
 
 
 
 
 
 
 
5.6. Calibration Error Code List