Teledyne MODEL T703 ユーザーズマニュアル

ページ / 287
Operating The T703 over the Serial I/O Ports 
Teledyne API T703 Calibrator Operation Manual 
108  
 
7.1.4. COM PORT COMMUNICATION MODES 
Each of the calibrator’s serial ports can be configured to operate in a number of different modes, listed in Table 
7-1.  As modes are selected, the calibrator sums the Mode ID numbers and displays this combined number on 
the front panel display.  For example, if quiet mode (01), computer mode (02) and Multi-Drop-enabled mode (32) 
are selected, the Calibrator would display a combined MODE ID of 35.   
Table 7-1:  COM Port Communication Modes 
MODE
1
 ID 
DESCRIPTION 
QUIET 
Quiet mode suppresses any feedback from the calibrator (such as warning messages) 
to the remote device and is typically used when the port is communicating with a 
computer program where such intermittent messages might cause communication 
problems.   
Such feedback is still available but a command must be issued to receive them. 
COMPUTER  
Computer mode inhibits echoing of typed characters and is used when the port is 
communicating with a computer operated control program. 
SECURITY  
When enabled, the serial port requires a password before it will respond.  The only 
command that is active is the help screen (?  CR).   
E, 7, 1 
2048 
When turned on this mode switches the COM port settings 
 from 
No parity; 8 data bits; 1 stop bit 
to 
Even parity; 7 data bits; 1 stop bit 
RS-485 
1024 
Configures the COM2 Port for RS-485 communication.  RS-485 mode has precedence 
over multidrop mode if both are enabled. 
MULTIDROP 
PROTOCOL 
32 
Multidrop protocol allows a multi-instrument configuration on a single communications 
channel.  Multidrop requires the use of instrument IDs. 
ENABLE 
MODEM 
64 
Enables to send a modem initialization string at power-up.  Asserts certain lines in the 
RS-232 port to enable the modem to communicate. 
ERROR 
CHECKING
128 
Fixes certain types of parity errors at certain Hessen protocol installations. 
XON/XOFF 
HANDSHAKE
2
 
256 
Disables XON/XOFF data flow control also known as software handshaking. 
HARDWARE 
HANDSHAKE 
Enables CTS/RTS style hardwired transmission handshaking.  This style of data 
transmission handshaking is commonly used with modems or terminal emulation 
protocols as well as by Teledyne Instrument’s APICOM software. 
HARDWARE 
FIFO
2
 
512 
Disables the HARDWARE FIFO (First In – First Out), When FIFO is enabled it 
improves data transfer rate for that COM port. 
COMMAND 
PROMPT 
4096 
Enables a command prompt when in terminal mode. 
1
 Modes are listed in the order in which they appear in the  
SETUP  MORE  COMM  COM[1 OR 2]  MODE menu 
2
 The default setting for this feature is ON.  Do not disable unless instructed to by Teledyne API’s Customer Service 
personnel. 
 
Note 
 Communication Modes for each COM port must be configured independently. 
07223B DCN6378