Teledyne API - T100 UV Fluorescence SO2 Analyzer
Getting Started
65
zero air produces higher concentrations of SO
2
. The T100 usually has a constant
flow rate and a constant permeation rate; hence, variations in concentration can be
achieved by changing the IZS temperature.
achieved by changing the IZS temperature.
3.3.2.7. EXTERNAL ZERO AIR SCRUBBER
The IZS option includes an external zero air scrubber assembly that removes all SO
2
from the zero air source. The scrubber is filled with activated charcoal.
3.3.2.8. PNEUMATIC LAYOUT WITH O
2
SENSOR OPTION
Figure 3-21 shows the internal, pneumatic connections for the analyzer with the oxygen
(O
(O
2
) sensor option installed. Pneumatically, the O
2
sensor draws a flow of 80 cm³/min in
addition to the normal sample flow rate. It is separately controlled with its own critical
flow orifice.
flow orifice.
FLOW PRESSURE
SENSOR PCA
Chassis
with O
2
Sensor Option
EXHAUST
GAS OUTLET
UV
LAMP
PMT
REACTION
CELL
HYDROCARBON
SCRUBBER
(Kicker)
VACUUM
MANIFOLD
O
2
Sensor
Flow Control
O
2
Sensor
SAMPLE
GAS INLET
PUMP
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Figure 3-21:
Pneumatic Layout with O
2
Sensor
06807C DCN6650