Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Getting Started 
 
 
65 
zero air produces higher concentrations of SO
2
. The T100 usually has a constant 
flow rate and a constant permeation rate; hence, variations in concentration can be 
achieved by changing the IZS temperature.  
3.3.2.7. EXTERNAL ZERO AIR SCRUBBER 
The IZS option includes an external zero air scrubber assembly that removes all SO
2
 
from the zero air source. The scrubber is filled with activated charcoal. 
3.3.2.8. PNEUMATIC LAYOUT WITH O
2
 SENSOR OPTION 
Figure 3-21 shows the internal, pneumatic connections for the analyzer with the oxygen 
(O
2
) sensor option installed. Pneumatically, the O
2
 sensor draws a flow of 80 cm³/min in 
addition to the normal sample flow rate.  It is separately controlled with its own critical 
flow orifice.   
FLOW PRESSURE 
SENSOR PCA
Chassis
with O
2
Sensor Option
EXHAUST 
GAS OUTLET
UV 
LAMP
PMT
REACTION 
CELL
HYDROCARBON 
SCRUBBER
(Kicker)
VACUUM
 MANIFOLD
O
2
Sensor
Flow Control
O
2
Sensor
SAMPLE 
GAS INLET
PUMP
SAMPLE 
PRESSURE 
SENSOR
FLOW 
SENSOR
 
Figure 3-21: 
Pneumatic Layout with O
2
 Sensor  
06807C DCN6650