Teledyne API - T100 UV Fluorescence SO2 Analyzer  
66 
3.3.2.9. PNEUMATIC LAYOUT WITH CO
2
 SENSOR OPTION 
Figure 3-22 shows the internal, pneumatic connections for the analyzer with the carbon 
dioxide (CO
2
) sensor option installed. Pneumatically, the CO
2
 sensor is placed in line 
with the sample gas line between the particulate filter and the analyzer’s sample 
chamber.  It does not alter the gas flow rate of the sample through the analyzer.  
 
FLOW PRESSURE 
SENSOR PCA
Chassis
with CO
2
Sensor Option
EXHAUST 
GAS OUTLET
UV 
LAMP
PMT
REACTION 
CELL
PUMP
HYDROCARBON 
SCRUBBER
(Kicker)
CO
2
Probe
SAMPLE 
GAS INLET
SAMPLE 
PRESSURE 
SENSOR
FLOW 
SENSOR
 
Figure 3-22: 
Pneumatic Layout with CO
2
 Sensor 
3.3.2.10. ABOUT ZERO AIR AND CALIBRATION (SPAN) GASES 
Zero air and span gas are required for accurate calibration. 
Z
ERO 
A
IR
 
A gas that is similar in chemical composition to the earth’s atmosphere but without the 
gas being measured by the analyzer, in this case SO
2
. If your analyzer is equipped with 
an Internal Zero Span (IZS) or an external zero air scrubber option, it is capable of 
creating zero air. 
For analyzers without an IZS or external zero air scrubber option, a zero air generator 
such as the Teledyne API Model 701 can be used (Figure 3-16)