31 
MAN-0076 Rev 05 Millennium II 
December 07, 2012 
Net Safety Monitoring Inc 
3.  Activate the enter key (switch 3) to display ‘slave address’ (default address: 001). 
 
4.  Use the up key (switch 1) to increase the address and the down key (switch 2) to decrease the value. The value 
range is 001-247. 
 
5.  Activate the enter key (switch 3) when the desired value is displayed.  
 
6.  After setting the Slave Address, exit to this sub menu option using switch 3.  
 
7.  Activate the down key (switch 2) to highlight ‘baud rate’, then activate the enter key (switch 3) to display the 
current baud rate.  
 
8.  Use the up key (switch 1) to increase the baud rate and the down key (switch 2) to decrease it.  
 
9.  Activate the enter key (switch 3) when the desired value is displayed.  
 
10.  After setting the baud rate, exit this sub menu option using switch 3, and then activate the down arrow key 
(switch 2) to highlight ‘Parity Bit’.  
 
11.  Activate switch 3, then activate the up key (switch 2), or the down key (switch 1) to choose a value.   
 
12.  Activate the exit key (switch 3) when the desired value is displayed, then select “Exit” at each menu stage 
(sub menu and main menu).   
  
4.3.0 Setup Current Date 
This option allows you to set the current date for event logging. The default date is set at the factory in Mountain Time 
(MT).  
 
1.  Enter the main menu by selecting/activating any key to get the “enter main menu” prompt, then activate 
switch 1 to select “yes”. 
 
2.  Activate the up key (switch 1) or down key (switch 2) until “Setup Current Date?” option is displayed.  
 
3.  Activate the enter key (switch 3) to display the sub menu option ‘year’, ‘month’, ‘day’
 
4.  Activate the up key (switch 1) to change the current year/month/day settings and switch 2 to cycle across 
year’, ‘month’, ‘day’ values and ‘OK’.  
 
13.  After desired setting are made, navigate to “OK?” and activate the enter key (switch 3) to confirm. To exit 
main menu, select “Exit” at each menu stage (sub menu and main menu).   
 
4.3.1 Setup Current Time 
This option allows you to set the current time for event logging. The default time is in Mountain Time (MT)  
 
1.  Enter the main menu by activating any key to get the “enter main menu” prompt, then activate switch 1 to 
select “yes”. 
 
2.  Activate the up key (switch 1) or down (switch 2) until “Setup Current Time?” option is displayed.