The area of measurement is not greater than a circle of diameter 0.1H.
Luminance uniformity on white
White point chromaticity and luminance is calibrated at the center of the display for x=0.3127,
y=0.3290 and Y=100 cd/m
y=0.3290 and Y=100 cd/m
2
(at level 1019).
The least bright point has a luminance which is equal to or larger than 88% of the luminance
measured in the center of the screen.
measured in the center of the screen.
Color uniformity on white
White point chromaticity and luminance is calibrated at the center of the display for x=0.3127,
y=0.3290 and Y=100 cd/m
y=0.3290 and Y=100 cd/m
2
(at level 1019).
For all points the following tolerance requirement applies:
Maximal:
The tolerances apply for a temperature of 25°C.
The tolerances apply for a temperature of 25°C.
Luminance uniformity on black
For all points, the measured luminance value deviates less than 50% from the luminance value
measured at the center point.
measured at the center point.
7.1.1.3 Power supply specification
7.1.1.3.1
Power inputs
• AC input voltage:
90 - 264 Vac
47 - 63 Hz
Input current: 3.6 A at 115 Vac and 1.8 A at 230 Vac
Inrush current: maximum 14 A at 230 Vac (cold start) / maximum 7 A at 120 Vac (@ 25 °C)
7.1.1.3.2
Power consumption
• Nominal power consumption: 100 W
• Maximum power consumption: 115 W
• Power consumption in hard standby mode: lower than 3 W
• Power consumption in soft standby mode: typical 65 W
• Maximum power consumption: 115 W
• Power consumption in hard standby mode: lower than 3 W
• Power consumption in soft standby mode: typical 65 W