3M 747 ユーザーズマニュアル

ページ / 30
13
The ALL setting is described in Section 4.1, “Open Collector Output.”
Error Message Meaning
Solution
Err.1
Finger removed from touch panel during 
measurement
Repeat the test
Err.2
Actual test type does not match specified 
test type
Check test type (location), and 
repeat the test
O.F.
Measured value exceeded 200.0 
MΩ 
(Overflow)
Check test type and connections, 
and repeat the test
3.1 ESD Shoe Testing with the 3M
 Shoes and Wrist Strap Tester 747
This procedure tests the effectiveness of ESD protective shoes, and displays the 
actual measured resistance value along with the qualitative test result.
1.  Ensure that the supplied shoes testing cable is connected between the shoe 
testing plate and the shoe testing connector. Refer to section 2.1, “Connecting 
the ESD Shoe Testing Plate.”
2.  Select the required threshold resistance values. Refer to Section 2.2, 
“Threshold Resistance Settings (1).”
3.  The wrist strap must not be connected to the front panel connector during 
shoe testing. If it is connected, remove it. 
4.  Step on the shoe testing plate, then press the touch panel with your finger. 
Continue pressing the touch panel until measurement is completed (until the                          
 indicator goes out).
 
Testing starts and the following indicators are displayed: 
 • 
 and 
 symbol appear on the LCD.
 
• The shoe LED lights yellow.