AACC 2000 Carbon                                                                       Nov.  1, 1997
86
User Calibration Enable
 
The User calibration facility must first be enabled in configuration level by setting the
parameter ‘
UCAL' in the input conf list to 'YES'.   This will make the User calibration
parameters visible in Operator ‘
FuLL’ level.
 
Select configuration level as shown in Chapter 5, Configuration.
 
   
 
 
 
+
 
 
 
 
 
 
 
The Calibration Configuration List
 
Press  
  until you reach the ‘
CAL-Conf’ list.
 
Press 
 until you reach ‘
UCAL’.
 
User Calibration Enable
 
Use 
 
 or  
 to select:
 
YES:
    Calibration enable
 
no:
      Calibration disabled
 
Press  
  and  
  together to go to the 
Exit
display.
 
 
Exit configuration
 
Use 
  or  
 
to select ‘
YES’ to return to Operator
level.