Marathon Computer AACC 2000 Manual Do Utilizador

Página de 103
Marathon Monitors Inc.
AACC 2000 Carbon                                                                       Nov.  1, 1997
86
User Calibration Enable
 
The User calibration facility must first be enabled in configuration level by setting the
parameter ‘
UCAL' in the input conf list to 'YES'.   This will make the User calibration
parameters visible in Operator ‘
FuLL’ level.
 
Select configuration level as shown in Chapter 5, Configuration.
 
   
 
 
 
+
 
 
 
 
 
 
 
The Calibration Configuration List
 
Press  
  until you reach the ‘
CAL-Conf’ list.
 
Press 
 until you reach ‘
UCAL’.
 
User Calibration Enable
 
Use 
 
 or  
 to select:
 
YES:
    Calibration enable
 
no:
      Calibration disabled
 
Press  
  and  
  together to go to the 
Exit
display.
 
 
Exit configuration
 
Use 
  or  
 
to select ‘
YES’ to return to Operator
level.