RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 12 
  USE THE CORRECT PRESSURE CONNECTORS:  The RPM4/HPMS TEST port fitting is a 
DH500 F (see Section 1.2.1).  It is NOT a 1/8 in. NPT F.  Never use a fitting other than the 
corresponding male fitting in these connectors.  Damage to the connectors and dangerous failure 
under pressure could result from using incorrect fittings. 
 
  DO NOT APPLY PRESSURE UNTIL YOU ARE FAMILIAR WITH OPERATION:  The RPM4/HPMS 
rear panel test port connects internally to both the RPM4’s 3 000 psi (20 MPa) and 10 000 psi 
(70 MPa) Q-RPTs.  The valve on the front of the HPMS isolates the low pressure Q-RPT when the 
high pressure Q-RPT is in use.  Do not apply pressure to the RPM4/HPMS until you are familiar 
with its operation and know how to protect the low pressure Q-RPT from overpressure (see 
Section 3.2.5).  FAILURE TO PROTECT THE LOW PRESSURE Q-RPT FROM OVERPRESSURE MAY 
DESTROY IT.  DAMAGE DUE TO Q-RPT OVERPRESSURE IS NOT COVERED BY THE PRODUCT 
WARRANTY. 
 
2.4.6.1 
THE ATM PORTS OF RPM4 
The  ATM ports on the RPM4 Q-RPT modules are connected to the RPM4’s 
internal barometer to assure that the RPM4 gauge pressure measurements are 
relative to ambient pressure.  These ports should always be left completely 
unobstructed and open to atmosphere. 
 
  NEVER plug, obstruct or connect a supply pressure to the RPM4 Q-RPT module 
ATM ports.  This may adversely affect GAUGE mode operation and AutoZeroing 
functions. 
 
2.4.7 
CHECK/SET SECURITY LEVEL 
RPM4 has a security system based on user levels.  By default, the security system is set to 
“low”, which includes access restriction to internal calibration coefficients, and there is no 
password required to change the security level.  See Section 3.5.5.5 for information on the 
security level system.  As part of the RPM4 startup, determine the security level that is 
appropriate for the RPM4 and set a password if desired. 
 
  RPM4 is delivered with the security level set to “low” to avoid inadvertent altering of critical 
internal settings but with access to changing security levels unrestricted.  It is recommended that the 
low security level be maintained at all times and password protection be implemented if control 
over setting of security levels is desired. 
 
2.4.8 
SETTING UP AUTOTEST FILES 
RPM4 supports automated test/calibration sequences.  AutoTest sequence 
parameters for testing specific DUTs can be stored in File AutoTest files and 
recalled to run a test.  Consider setting up File AutoTest files for frequently tested 
DUTs as part of the RPM4 set up process (see Section 3.3.10, 3.4.5.2).