3. OPERATION 
 
Page 39 
© 2007 DH Instruments, a Fluke Company 
  When a head correction is being applied, it is indicated by <h> in the fourth character from 
the right in the top line of the MAIN RUN screen (see Section 3.1.1).  When the head correction is 
zero, the <h> is not shown. 
 
  The default head height unit is inches and the fluid species is Nitrogen (N
2
).To change units 
of head height between inches and centimeters and to change the test fluid species, use 
[SPECIAL], <3head> (see Section 3.5.3). 
 
3.3.8 
[SDS] 
 PURPOSE 
The SDS function is not used in RPM4 A70M/A20M-AF. 
3.3.9 
 [A
UTO
Z] 
  PURPOSE 
To run the AutoZ function that rezeros the active Q-RPT between full calibrations. 
 
  See Section 3.5.1,  Principle for a full explanation of the AutoZ function. 
 
  PRINCIPLE 
Run AutoZ is the function by which the current Q-RPT reading is compared to a reference, 
P
std,0
, at atmospheric pressure to determine a new value of P
offset
.  The value of P
offset
 is then 
used by AutoZ to automatically correct the Q-RPT for possible changes in zero over time 
(see Section 3.5.1, 
 PRINCIPLE). 
3.3.9.1 
[AUTOZ] IN GAUGE MODE 
  PURPOSE 
To “rezero” the active Q-RPT in gauge measurement mode. 
 
   See Section 3.51,  Principle for a full explanation of the AutoZ function. 
 
  OPERATION 
 
  For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the 
active range and measurement mode.  AutoZ ON is indicated by <z> to the left of the 
Q-RPT designator on the first line of the main run screen.  AutoZ ON and OFF is set 
using [SPECIAL], <1AutoZ> (see Section 3.51).  If AutoZ is OFF for the active Q-RPT, 
and measurement mode, <AutoZ is off> is displayed when [AutoZ] is pressed. 
 
To run AutoZ in gauge measurement mode, use [MODE] to set gauge as the 
measurement mode (see Section 3.3.3) and press [AutoZ] from the main run 
screen.  <Running gauge AutoZ> is displayed briefly before returning to the 
main run screen.