RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 40 
  Before running AutoZ in gauge mode, ensure that the pressure applied to the 
Q-RPT is truly zero gauge (atmospheric pressure).  If running AutoZ in gauge mode 
results in a zero offset that RPM4 considers unusually large, <Confirm 0 gauge P!> is 
displayed.  Check that zero gauge pressure is applied to the Q-RPT module rear panel 
TEST(+) port and the ATM port is open to atmosphere and unobstructed. 
 
  Allow the RPM4 to stabilize at atmospheric pressure and ambient temperature 
for 2 to 3 minutes before running AutoZ  in gauge mode. 
 
3.3.9.2 
[AUTOZ] IN ABSOLUTE MODE 
  PURPOSE  
To “rezero” the active Q-RPT in absolute measurement mode 
 
  See Section 3.5.1,  Principle for a full explanation of the AutoZ function. 
 
  OPERATION 
 
  For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the 
active Q-RPT and measurement mode.  AutoZ ON is indicated by <z> to the left of 
the range designator on the first line of the main run screen.  AutoZ ON and OFF is 
set using [SPECIAL], <1AutoZ> (see Section 3.5.1).  If AutoZ is OFF for the active 
range, and measurement mode, <AutoZ OFF> is displayed when [AutoZ] is pressed. 
 
To run AutoZ in absolute measurement mode, use [MODE] to set absolute 
measurement mode (see Section 3.3.3) and press [AutoZ] from the main run 
screen.  The display is: 
1. 
Active Q-RPT designator. 
2. 
Current measurement mode (absolute). 
3. 
Selection of source of P
std,0
 reference to which 
to AutoZ. 
       
     
 
 AutoZ Hi abs by: 
13ntry 2COM2 3Lo RPT 
                  
                     
 
Selecting <1entry> allows the value of P
std,0
 to be entered from the front panel keypad.  
Selecting <2COM2> allows the value of P
std,0
 to be read automatically from another 
DHI RPMx connected to RPM4’s COM2 communications port. 
Selecting  <3Lo RPT > is shown only if the Hi Q-RPT is active.  This selection 
allows the value of P
std,0
 for the Hi Q-RPT to be read automatically from the 
RPM4s Lo Q-RPT. 
 
  Allow the RPM4 to stabilize at atmospheric pressure and ambient temperature 
for 5 to 10 minutes before running AutoZ in absolute mode.