RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 70 
Table 9.  Reset – Sets 
RESET 
RESULT     
SEE SECTION 
[UNIT] 
Pressure unit of measure to first of the six available (psi) 
3.3.2 
[MODE] 
Measurement mode to the native mode of  the Q-RPT (absolute) 
3.3.3 
[RANGE] 
Hi Q-RPT default range (10000 psi) 
3.3.1 
[HEAD] 
0 cm height and Nitrogen medium 
3.3.7 
Stability Limit 
0.005 % of Hi Q-RPT default range 
3.4.3 
Upper Limit 
Hi Q-RPT default range default value 
3.4.4 
Resolution 
0.001 % FS of Hi Q-RPT default range 
3.4.2 
AutoZ 
AutoZ ON for both Q-RPTs and gauge and asbsolute 
measurement modes.   Z
offset
 value not affected 
3.5.1 
AutoZ 
P
atm,0
 set to 101.325 kPa a 
3.5.1, 3.2.2 
Leak Check 
15 second run time.  Clear logged results 
3.3.5 
Screen Saver 
10 minutes to activation 
3.5.5.1 
Key Sounds 
Medium tone valid key press sound 
3.5.5.2 
Lo Vnt 
Automatic 
3.5.7.4 
ReadRate Automatic 
3.5.7.2 
3.5.9.2 
<2 UNITS> 
 PURPOSE 
Sets the pressure units available under the UNIT function to default selections 
depending (see Section 3.3.2). 
Sets the user defined unit to 1.000/Pa (see Section 3.5.6). 
Sets the reference temperature for inWa unit to 20°C. 
3.5.9.3 
<3AT
EST
 PURPOSE 
•  Resets Auto Tests to default characteristics (see Section 3.3.10).  
•  Clears File AutoTest definition files (see Section 3.4.5.2). 
•  Clears AutoTest results data logs (see Section 3.4.5.1).