3. OPERATION 
 
Page 71 
© 2007 DH Instruments, a Fluke Company 
3.5.9.4 
<4 CAL> 
 PURPOSE 
 
  The Reset - Cal function will reset the Q-RPT and barometer calibration 
coefficients and reset AutoZ values to zero.  This will change the RPM4 calibration 
and could cause it to make out of tolerance measurements. 
 
Clears all user values affecting the calibration of Q-RPTs and the on-board 
barometer.  The Reset – Cal resets are itemized in Table 10. 
Table 10.  Reset – Cal 
RESET 
RESULT     
SEE SECTION 
All Q-RPT calibration 
coefficients 
PA to zero, PM to 1 
5.3.1 
Q-RPT absolute mode 
ON 
5.3.5 
On-board barometer 
calibration coefficients 
PA to zero, PM  to 1 
5.2.2 
Calibration date 
Set all dates to 19800101 
5.3.7 
AutoZ values 
All  P
offset
 values to zero for absolute mode and 
101325 Pa for gauge and negative gauge modes 
3.5.1 
AutoZ function 
ON, all Q-RPT s, all measurement modes 
3.5.1 
3.5.9.5 
<5 ALL> 
 PURPOSE 
 
  The reset - all function clears and deletes large amounts of user defined 
information including critical calibration data. 
 
Combines all resets in one global reset command that clears the entire user 
section of non-volatile memory except the ID function (see Section 3.5.5.4) and 
the security level password (see Section 3.5.5.5) returning RPM4 to the  
“as delivered” condition.  The Reset – All resets are itemized in Table 11. 
Table 11.  Reset – All 
RESET 
RESULT 
SEE SECTION 
Reset – Sets 
All the resets of Reset - Sets 
3.5.9.1 
Reset – Units 
All the resets of Reset - Units 
3.5.9.2 
Reset – Atest 
All the resets of Reset - ATest 
3.5.9.3 
Reset – Cal 
All the resets of Reset - Cal 
3.5.9.4 
Remote Interfaces 
COM1, COM2 and IEEE-R88 interfaces to 
default settings. 
3.5.2 
Remote Communications 
Remote command format to Classic 
3.5.2.3 
Level (Security) 
Reset security level to low 
3.5.5.5 
Parallel Measurement Mode 
Off 
3.5.7.3