Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
2-14
 
Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Setting Measurement Parameters for the Power Meter Calibration
1. Connect the measurement equipment as shown in Step 1 of 
Figure 2-11  Connections for R Channel and Source Calibration
CAUTION
Note that the front panel jumper between R In and R Out must remain 
installed during the procedure steps that use the connections in Step 1 of 
2. From the front panel of the analyzer, set the desired receiver (RF) frequency and source 
output power by pressing:
 
 
  
  
 
   
  
Note that these are the example RF start and stop frequencies. Enter the 
RF start and stop frequencies for your measurement instead.
If the LO frequency is not set to 0 Hz, press:
 
 
3. To select the measurement trace, press:
 
 
The measurement trace is shown on the display.
4. Select the analyzer as the system controller:
 
System
 INSTRUMENT MODE   FREQ OFFS MENU 
Start
650
M/
µ
Stop
950
M/
µ
 LO FREQUENCY 
0
x1
Meas
 INPUT PORTS   R 
Local
 SYSTEM CONTROLLER