Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Setting the Analyzer to Make an R Channel Measurement
1. Connect the equipment as shown in 
.
Figure 2-12  R-Channel Mixer Measurement Equipment Setup
NOTE
An error message will be displayed while the R In port is disconnected. 
Ignore this error message until step 3 is complete.
The analyzer is now displaying the conversion loss of the mixer calibrated with power 
meter accuracy.
2. While the analyzer is still set to the IF frequency range, set the frequency offset mode 
LO frequency from the analyzer by pressing:
 
 
 
 
Note that this is the example LO frequency. Enter the LO frequency for 
your measurement instead.
The LO menu is used to set only the LO CW frequency. All other settings apply when 
using the HP/Agilent 8625A external source.
3. To select the converter type and a high-side LO measurement configuration, press:
 
 
Note that these are the example settings. Enter the settings for your 
measurement instead.
System
 INSTRUMENT MODE   FREQ OFFS MENU 
 LO FREQUENCY 
1000
M/
µ
 RETURN   DOWN CONVERTER   RF < LO