Making Mixer Measurements (Option 089 Only)
High Dynamic Range Swept RF/IF Conversion Loss
NOTE
Because power meter calibration requires a longer sweep time, you may want 
to reduce the number of points before pressing 
. After the 
power meter calibration is finished, return the number of points to its original 
value and the analyzer will automatically interpolate this calibration.
Perform a Receiver Calibration Over the IF Range
1. Connect the measurement equipment as shown in 
.
Figure 2-17 Connections for Receiver Calibration
2. To calibrate the B-channel over the IF range, press:
 
 
 
 
 
 
 (
 
, 8722ES) 
Once completed, the analyzer should display 
5 dBm (
10 dBm, 8722ES).
Set the Analyzer to the RF Frequency Range
You can find the RF frequency range by using a simple calculation or using the mixer 
measurement diagram on the analyzer display.
Using the Calculation
Add the LO frequency to the IF frequency start and stop values. If using an LO frequency 
of 1500 MHz, the RF start frequency would be 1.6 GHz (1500 + 100 MHz) and the stop 
frequency would be 2.5 GHz (1500 + 1000 MHz).
 TAKE CAL SWEEP 
Meas
 INPUT PORTS   B 
Cal
 CALIBRATE MENU   RECEIVER CAL 
5
x1
10
x1
 TAKE RCVR CAL SWEEP