Making Mixer Measurements (Option 089 Only)
Isolation Example Measurements
7. To adjust the display scale, press:
 
The measurement results show the mixer’s LO to RF isolation.
Figure 2-35 Example Mixer LO to RF Isolation Measurement 
RF Feedthrough 
The procedure and equipment configuration necessary for this measurement are very 
similar to those of the previous LO to RF Isolation procedure, with the addition of an 
external source to drive the mixer’s LO port as we measure the mixer’s RF feedthrough. 
RF feedthrough measurements do not use the frequency offset mode.
1. Select the CW LO frequency and source power from the front panel of the external 
source.
CW frequency = 300 MHz
Power = 10 dBm
2. Initialize the analyzer by pressing 
.
3. To select the analyzer frequency range and source power, press:
 
 
 
 
 
 
 
 (
 
, 8722ES)
This source stimulates the mixer’s LO port.
4. To select a ratio B/R measurement, press:
 
 
Scale Ref
 AUTO SCALE 
Preset
Start
10
M/
µ
Stop
3000
M/
µ
Power
0
 PWR RANGE MAN 
x1
10
x1
Meas
 INPUT PORTS   B/R