Making Mixer Measurements (Option 089 Only)
Isolation Example Measurements
Isolation Example Measurements 
Isolation is the measure of signal leakage in a mixer. Feedthrough is specifically the 
forward signal leakage to the IF port. High isolation means that the amount of leakage or 
feedthrough between the mixer’s ports is very small. Isolation measurements do not use 
the frequency offset mode
 illustrates the signal flow in a mixer.
Figure 2-32 Signal Flow in a Mixer Example 
LO to RF Isolation 
LO to RF isolation is the amount the LO power is attenuated when it appears directly at 
the RF port.
1. Initialize the analyzer by pressing 
.
2. To select the analyzer frequency range and source power, press:
 
 
 
 
 
 
 
 (
 
, 8722ES)
This source stimulates the mixer’s LO port.
3. To select a ratio B/R measurement, press:
 
 
Preset
Start
10
M/
µ
Stop
3000
M/
µ
Power
0
 PWR RANGE MAN 
x1
10
x1
Meas
 INPUT PORTS   B/R