Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
3- 5
Making Time Domain Measurements
Making Transmission Response Measurements
Making Transmission Response Measurements 
In this example measurement there are three components of the transmission response:
• RF leakage at near zero time
• the main travel path through the device (1.6 
µ
s travel time)
• the "triple travel" path (4.8 
µ
s travel time) 
This example procedure also shows you how time domain analysis allows you to 
mathematically remove individual parts of the time domain response to see the effect of 
potential design changes. This is accomplished by "gating" out the undesirable responses. 
With the "gating" capability, the analyzer time domain allows you to perform "what if" 
analysis by mathematically removing selected reflections and seeing the effect in the 
frequency domain.
1. Connect the device as shown in 
.
Figure 3-2 Device Connections for Time Domain Transmission Example 
Measurement 
2. To choose the measurement parameters, press:
 
 or on ET models: 
 
 
 
 
 
3. Substitute a thru for the device under test and perform a frequency response correction. 
Refer to 
4. Reconnect your device under test.
Preset
Meas
 Trans: FWD S21 (B/R) 
 TRANSMISSN 
Start
119
M/
µ
Stop
149
M/
µ
Scale Ref
 AUTO SCALE