3- 9
Making Time Domain Measurements
Making Reflection Response Measurements
Making Reflection Response Measurements
The time domain response of a reflection measurement is often compared with the time 
domain reflectometry (TDR) measurements. Like the TDR, the analyzer measures the size 
of the reflections versus time (or distance). Unlike the TDR, the time domain capability of 
the analyzer allows you to choose the frequency range over which you would like to make 
the measurement.
1. To choose the measurement parameters, press:
 
 or on ET models: 
 
 
 
 
2. Perform an S
11
 1-port correction on PORT 1. Refer to 
 for a detailed procedure.
3. Connect your device under test as shown in 
.
Figure 3-7 Device Connections for Reflection Time Domain Example 
Measurement 
4. To better view the measurement trace, press:
 
 shows the frequency domain reflection response of the cables under test. The 
complex ripple pattern is caused by reflections from the adapters interacting with each 
other. By transforming this data to the time domain, you can determine the magnitude 
of the reflections versus distance along the cable.
Preset
Meas
 Refl: FWD S11 (A/R) 
 TRANSMISSN 
Start
50
M/
µ
Stop
3
G/n
Scale Ref
 AUTO SCALE