Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-14
 
Making Measurements
Using Display Functions
However, there are two configurations that will not sweep continuously.
1. For analyzers with source attenuators, with channel 1 having one attenuation value 
and channel 2 set to a different attenuation value, then continuous sweep is disabled to 
avoid wear on the attenuator. A similar situation where this occurs is when a 2-port cal 
is active and the port 1 attenuation value is not equal to the port 2 attenuation value. 
Since one attenuator is used for both measurements, this would cause the attenuator to 
continuously switch power ranges, so continuous sweep is not allowed. (The exception is 
analyzers configured with option 400. Option 400 analyzers have two attenuators and 
can allow different attenuation settings on each port.)
2. For ES analyzers with Option 007 or 085 (options using a mechanical transfer switch), 
channel 1 is driving one test port and channel 2 is driving the other test port. This 
would cause the test port transfer switch to continually cycle. The instrument will not 
allow the transfer switch or attenuator to continuously switch ranges in order to update 
these measurements without the direct intervention of the operator.
If one of these conditions exist, the test set hold mode will engage, and the status notation 
tsH
 will appear on the left side of the screen. The hold mode leaves the measurement 
function in only one of the two measurements. To update both measurement setups, press 
 
. Refer to 
Viewing Four Measurement Channels
Four measurement channels can be viewed simultaneously by enabling auxiliary channels 
3 and 4. Although independent of other channels in most variables, channels 3 and 4 are 
permanently coupled to channels 1 and 2 respectively by stimulus. That is, if channel 1 is 
set for a center frequency of 200 MHz and a span of 50 MHz, channel 3 will have the same 
stimulus values.
NOTE
Channels 1 and 2 are referred to as primary channels and channels 3 and 4 
are referred to as auxiliary channels.
Channel 3 or 4 are activated when the Chan 3 or Chan 4 keys are pressed. Alternatively, 
you can enable the auxiliary setting 
 to ON. For example, if channel 1 is 
active, pressing 
 to ON enables channel 3 and its trace appears on the display. 
Channel 4 is similarly enabled and viewed when channel 2 is active.
1. Press 
 to select the type of display of the data. This example uses the log mag 
format.
2. If channel 1 is not active, make it active by pressing 
.
3. Press 
 
, set 
 to ON, set 
 to 
ON, and set 
 to 
.
The display will appear as shown in 
. Channel 1 is in the upper-left quadrant 
of the display, channel 2 is in the upper-right quadrant, and channel 3 is in the lower 
half of the display.
Sweep Setup
 MEASURE RESTART 
 AUX CHAN 
 AUX CHAN 
Format
Chan 1
Display
 DUAL | QUAD SETUP 
 DUAL CHAN 
 AUX CHAN 
 SPLIT DISP 
 4X