7- 13
Operating Concepts
Source Attenuator Switch Protection
Source Attenuator Switch Protection 
The programmable step attenuator of the source can be switched between port 1 and port 2 
when the test port power is uncoupled, or between channel 1 and channel 2 when the 
channel power is uncoupled. To avoid premature wear of the attenuator, measurement 
configurations requiring continuous switching between different power ranges are not 
allowed.
For example, continuous switching would be required if channels 1 and 2 of the analyzer 
are decoupled, power levels in two different ranges are selected for each channel, and dual 
channel display is engaged. To prevent continuous switching between the two power 
ranges, the analyzer automatically engages the test set hold mode after measuring both 
channels once. The active channel continues to be updated each sweep while the inactive 
channel is placed in the hold mode. (The status annotation 
tsH
 appears on the left side of 
the display.) If averaging is on, the test set hold mode does not engage until the specified 
number of sweeps is completed. The 
 and 
 
softkeys can override this protection feature.
Allowing Repetitive Switching of the Attenuator 
The 
 and 
 softkeys allow measurements 
which demand repetitive switching of the step attenuator. Use these softkeys with caution; 
repetitive switching can cause premature wearing of the attenuator.
 causes one measurement to occur before activating the test set 
hold mode.
 causes a specified number of measurements to occur before 
activating the test set hold mode.
MEASURE RESTART 
NUMBER OF GROUPS 
MEASURE RESTART 
NUMBER OF GROUPS 
MEASURE RESTART 
NUMBER OF GROUPS