Optimizing Measurement Results
Increasing Sweep Speed
Sweep Speed-Related Errors 
IF delay occurs during swept measurements when the signal from the analyzer source is 
delayed in reaching the analyzer receiver because of an electrically long device. The 
receiver has a narrow IF band pass filter that tracks the receiver frequency because the 
receiver is sweeping. The delayed signal will be attenuated because the center of the 
internal IF filter has moved.
For most measurements, swept list mode will be the optimum choice. If there is any doubt 
about the effect of IF delay, perform the following test:
1. Set up the measurement using the swept list mode, as in the previous procedure.
2. Make the measurement and save the data trace to memory:
  
 
and 
3. Then switch to stepped list mode:
 
 
 
 
• If there is no difference between the measurements in either list mode, then use the 
swept list mode.
• If the memory trace indicates that there is more attenuation in swept list mode, it 
may be due to IF delay. You can usually remedy this problem by increasing the sweep 
time.
NOTE
IF bandwidths of 30 to 10 Hz cause the sweep (or that segment of the sweep) 
to always be stepped, thus eliminating IF delay.
To Decrease the Frequency Span 
The hardware of the network analyzer sweeps the frequency range in separate bands, 
where switching from band to band takes time. Modify the frequency span to eliminate as 
many band switches as possible while maintaining measurement integrity. Refer to the 
following table to identify the analyzer's band switch points:
Table 5-3  Band Switch Points 
Band 
Frequency Span 
Band 
Frequency Span 
1
50 MHz to 110 MHz
8
2.55 GHz to 4.71 GHz
2
110 MHz to 230 MHz
9
4.71 GHz to 8.256 GHz
3
230 MHz to 470 MHz
10
8.256 GHz to 13.562 GHz
4
470 MHz to 698 MHz
11 (8720ET/ES 
and 8722ET/ES)
13.562 GHz to 20.05 GHz
5
698 MHz to 1.17 GHz
12 (8722ET/ES)
20.05 GHz to 25.0 GHz
6
1.17 GHz to 1.878 GHz
13 (8722ET/ES)
25.0 GHz to 40.0 GHz
7
1.878 GHz to 2.55 GHz
Display
 DATA 
 MEMORY   DISPLAY: DATA 
 MEMORY 
Sweep Setup
 SWEEP TYPE MENU   EDIT LIST   LIST TYPE: [STEPPED]   DONE