Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
5-12
 
Optimizing Measurement Results
Increasing Sweep Speed
To Set the Auto Sweep Time Mode 
Auto sweep time mode is the default mode (the preset mode). This mode maintains the 
fastest sweep speed possible for the current measurement settings.
• Press 
 
 
, to re-enter the auto mode.
To Widen the System Bandwidth 
1. Press 
 .
2. Increase the IF bandwidth to increase the sweep speed.
The specifications and characteristics chapter of the reference guide shows the relative 
increase in sweep time as you decrease system bandwidth.
To Reduce the Averaging Factor 
By reducing the averaging factor (number of sweeps) or switching averaging off, you can 
increase the analyzer's measurement speed. The time needed to compute averages can also 
slow the sweep time slightly, in narrow spans.
1. Press 
 .
2. Enter an averaging factor that is less than the value displayed on the analyzer screen 
and press 
.
3. If you want to switch off averaging, press 
 
.
To Reduce the Number of Measurement Points 
1. Press 
 .
2. Enter a number of points that is less than the value displayed on the analyzer screen 
and press 
.
Refer to the “Specifications and Characteristics” chapter of the reference guide for 
examples of how sweep time changes with the number of points.
To Set the Sweep Type 
Different sweep speeds are associated with the following three types of non-power sweeps. 
Choose the sweep type that is most appropriate for your application.
1. Press 
 .
2. Select the sweep type:
• Select 
 for the fastest sweep for a given number of fixed points.
• Select 
 for the fastest sweep when specific non-linearly spaced 
frequency points are of interest.
• Select 
 for the fastest sweep when the frequency points of interest are in 
the lower part of the frequency span selected.
Sweep Setup
 SWEEP TIME 
0
Avg
 IF BW 
Avg
 AVG FACTOR 
x1
Avg
 AVERAGING OFF 
Menu
 NUMBER OF POINTS 
x1
Sweep Setup
 SWEEP TYPE MENU 
 LIN FREQ 
 LIST FREQ 
 LOG FREQ