Calibrating for Increased Measurement Accuracy
Calibration Considerations
Calibration Considerations 
Measurement Parameters 
Calibration procedures are parameter-specific, rather than channel-specific. When a 
parameter is selected, the instrument checks the available calibration data, and uses the 
data found for that parameter. For example, if a transmission response calibration is 
performed for B/R, and an S
11
 1-port calibration for A/R, the analyzer retains both 
calibration sets and corrects whichever parameter is displayed. Once a calibration has 
been performed for a specific parameter or input, measurements of that parameter remain 
calibrated in either channel, as long as stimulus values are coupled. In the response and 
response and isolation calibrations, the parameter must be selected before calibration. 
Other correction procedures select parameters automatically. Changing channels during a 
calibration procedure invalidates the part of the procedure already performed.
Device Measurements 
In calibration procedures that require measurement of several different devices, for 
example a short, an open, and a load, the order in which the devices are measured is not 
critical. Any standard can be re-measured, until the 
 key is pressed. The change in 
trace during measurement of a standard is normal.
“Response” and “response and isolation” calibrations require measurement of only one 
standard device. If more than one device is measured, only the data for the last device is 
retained.
Clarifying Type-N Connector Sex
When you are performing error correction for a system that has type-N port connectors, 
the softkey menus label the sex of the test port connector — not the calibration standard 
connector. For example, the label, SHORT (F), refers to the short that will be connected to 
the female test port. Since many devices have type-N (f) connectors, the calibration 
standard to select is the type-N (m). Be sure to use the port extension pin on the type-N 
(m) Open calibration standard.
NOTE
Since the 85032F calibration kit offsets are equal for both male and female 
connectors, the standard’s sex type is not requested during the calibration.
Omitting Isolation Calibration 
Isolation calibration can be omitted for most measurements, except where high dynamic 
range is a consideration. Use the following guidelines. When the measurement requires a 
dynamic range of:
• Less than 90 dB: Omit isolation calibration for most measurements.