Calibrating for Increased Measurement Accuracy
Calibration Considerations
• 90 to 100 dB: Isolation calibration is recommended with test port power greater than 
0 dBm. For this isolation calibration, averaging should be turned on with an averaging 
factor at least four times the measurement averaging factor. For example, use an 
averaging factor of 16 for the isolation calibration, and then reduce the averaging factor 
to four for the measurement after calibration.
• Greater than 100 dB: Same as 90 to 100 dB, but alternate mode should be used. See 
.
Saving Calibration Data 
You should save the calibration data, either in the internal non-volatile memory or on a 
disk. If you do not save it, it will be lost if you select another calibration procedure for the 
same channel, or if you change stimulus values. Instrument preset, power on, and 
instrument state recall will also clear the calibration data.
Restarting a Calibration 
If you interrupt a calibration to go to another menu, such as averaging, you can continue 
the calibration by pressing the 
 softkey in the correction 
menu.
The Calibration Standards 
During measurement calibration, the analyzer measures actual, well-defined standards 
and mathematically compares the results with ideal "models" of those standards. The 
differences are separated into error terms which are later removed during error correction. 
Most of the differences are due to systematic errors—repeatable errors introduced by the 
analyzer, test set, and cables—which are correctable.
The standard devices required for system calibration are available in compatible 
calibration kits with different connector types. Each kit contains at least one short circuit, 
one open circuit, and an impedance-matched load. In kits that require adapters for 
interface to the test set ports, the adapters are phase-matched for calibration prior to 
measurement of non-insertable and non-reversible devices. Other standard devices can be 
used by specifying their characteristics in a user-defined kit, as described in 
.
The accuracy improvement of the correction is limited by the quality of the standard 
devices, and by the connection techniques used. For maximum accuracy, ensure that the 
connectors are clean and use a torque wrench for final connections.
Electronic calibration (ECal) modules serve as the calibration standards for electronic 
calibration procedures. ECal modules are electronic networks that simulate impedance 
states that are similar to mechanical standards. The factory error correction for optimum 
performance is embedded in the module’s memory.
 RESUME CAL SEQUENCE