6- 19
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
12.Return the averaging to the original state of the measurement. For example, reduce the 
averaging factor by at least four times or turn averaging off.
13.To compute the isolation error coefficients, press:
 
 
 
The analyzer displays the corrected data trace. The analyzer also shows the notation 
Cor
 at the left of the screen, indicating that the correction is switched on for this 
channel.
NOTE
You can save or store the measurement correction to use for later 
measurements. Refer to 
 for procedures.
14.This completes the response and isolation correction for transmission measurements. 
You can connect and measure your device under test.
Response and Isolation Error Correction for Reflection 
Measurements 
The response and isolation error correction for reflection measurements provides the 
following benefits:
• removes frequency response of the test setup
• removes isolation in transmission measurements
• removes directivity in reflection measurements
To perform the response and isolation error correction for reflection measurements:
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, 
S
11
), leave the instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the 
reverse direction, S
22
), press:
 
3. Set any other measurement parameters that you want for the device measurement: 
power, sweep type, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
 RESUME CAL SEQUENCE   DONE   RESP ISOL’N CAL 
Preset
Meas
 Refl: REV S22 (B/R) 
Cal