Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
6- 17
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
Frequency Response and Isolation Error Corrections 
You can make a response and isolation correction for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements
NOTE
Although you can perform a response and isolation correction for reflection 
measurements, we recommend that you perform an S
11
 one-port error 
correction; it is more accurate and just as convenient.
Response and Isolation Error Correction for Transmission 
Measurements 
This procedure is intended for measurements that have a measurement range of greater 
than 90 dB.
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make a transmission measurement in the forward direction (S
21
), 
press:
 
 or on ET models: 
❏ For ES analyzers, if you want to make a transmission measurement in the reverse 
direction (S
12
), press:
 
3. Set any other measurement parameters that you want for the device measurement: 
power, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
5. If your calibration kit is different than the kit specified under the 
 softkey, 
press:
 
 (select your type of kit) 
If your type of calibration kit is not listed in the displayed menu, refer to 
.
6. To select a response and isolation correction and to start the response portion of the 
calibration, press:
 
 
Preset
Meas
 Trans: FWD S21 (B/R) 
 TRANSMISSN 
Meas
 Trans: REV S12 (A/R) 
Cal
 CAL KIT [ ] 
 CAL KIT   SELECT CAL KIT 
 RETURN 
 CALIBRATE MENU   RESPONSE & ISOL’N   RESPONSE