Agilent Technologies 20ET ユーザーズマニュアル
6- 25
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
b. Activate at least four times more averages than desired during the device
measurement.
c. Press
.
d. Return the averaging to the original state of the measurement, and press
.
18.To compute the error coefficients, press
.
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation
notation
Cor
at the left of the screen, indicating that error correction is on.
Enhanced Reflection Calibration
19.If you are measuring a bilateral device and want to remove the load match error,
activate the enhanced reflection calibration by pressing:
until ON is selected
NOTE
for procedures.
20.This completes the enhanced response correction procedure. You can connect and
measure your device under test.
Cal
RESUME CAL SEQUENCE ISOLATION FWD or REV
ISOL’N STD DONE
Cal
RESUME CAL SEQUENCE
DONE ENH RESP CAL
ENHANCED RESPONSE
ENH. REFL. on OFF