Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
b. Activate at least four times more averages than desired during the device 
measurement.
c. Press 
  
 
.
d. Return the averaging to the original state of the measurement, and press 
 
.
18.To compute the error coefficients, press 
.
The analyzer displays the corrected measurement trace. The analyzer also shows the 
notation 
Cor
 at the left of the screen, indicating that error correction is on.
Enhanced Reflection Calibration
19.If you are measuring a bilateral device and want to remove the load match error, 
activate the enhanced reflection calibration by pressing: 
 until ON is selected
NOTE
You can save or store the measurement correction to use for later 
measurements. Refer to 
 for procedures.
20.This completes the enhanced response correction procedure. You can connect and 
measure your device under test.
Cal
 RESUME CAL SEQUENCE   ISOLATION FWD or REV 
 ISOL’N STD DONE 
Cal
 RESUME CAL SEQUENCE 
 DONE ENH RESP CAL 
 ENHANCED RESPONSE 
 ENH. REFL. on OFF