Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
One-Port Reflection Error Correction 
• removes directivity errors of the test setup
• removes source match errors of the test setup
• removes frequency response of the test setup
You can perform a 1-port correction for an S
11
 or an S
22
 measurement for ES analyzers. 
The only difference between the two procedures is the measurement parameter that you 
select.
NOTE
This is the recommended error correction process for all reflection 
measurements, when full two-port correction or enhanced response 
calibration is not used.
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, 
S
11
), leave the instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the 
reverse direction, S
22
), press:
 
3. Set any other measurement parameters that you want for the device measurement: 
power, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
5. If your calibration kit is different than the kit specified under the 
 softkey, 
press:
 
 (select your type of kit) 
If your type of calibration kit is not listed in the displayed menu, refer to 
6. To select the correction type, press 
 and select the correction type.
❏ If you want to make a reflection measurement at PORT 1, press:
❏ For ES analyzers, if you want to make a reflection measurement at PORT 2, press:
7. Connect a shielded open circuit to PORT 1 (or PORT 2 for an S
22
 measurement).
Preset
Meas
 Refl: REV S22 (B/R) 
Cal
 CAL KIT [ ] 
 CAL KIT   SELECT CAL KIT 
 RETURN 
 CALIBRATE MENU 
 S11 1-PORT 
 S22 1-PORT