Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
Interpolation in Power Meter Calibration 
If the frequency is changed in linear sweep, or the start/stop power is changed in power 
sweep, then the calibration data is interpolated for the new range.
If calibration power is changed in any of the sweep types, the values in the power setting 
array are increased or decreased to reflect the new power level. Some accuracy is lost when 
this occurs.
Entering the Power Sensor Calibration Data 
Entering the power sensor calibration data compensates for the frequency response of the 
power sensor, thus ensuring the accuracy of power meter calibration.
1. Make sure that your analyzer and power meter are configured. Refer to the “Options 
and Accessories” chapter of the reference guide for configuration procedures.
2. Press 
  
 
.
The analyzer shows the notation 
EMPTY
, if you have not entered any segment 
information.
3. To create the first segment, press:
 
4. Enter the frequency of a correction factor data point, as listed on the power sensor, 
followed by the appropriate key: 
 
 
.
5. Press 
 and enter the correction factor that corresponds to the frequency 
that you have entered in the previous step. Complete the correction factor entry by 
pressing  
.
6. Repeat the previous three steps to enter up to 55 frequency segments.
You may enter multiple segments in any order because the analyzer automatically sorts 
them and lists them on the display by frequency value. The analyzer also automatically 
interpolates the values between correction factor data points.
If you only enter one frequency segment, the analyzer assumes that the single value is 
valid over the entire frequency range of the correction.
7. After you have entered all the frequency segments, press 
.
Editing Frequency Segments 
1. Access the Segment Modify menu by pressing 
 
 
 
 (or 
depending on where the segment is that you want to edit).
2. Identify the segment that you want to edit by pressing 
 and using the 
 
and 
 keys to locate and position the segment next to the pointer (>), shown on the 
display. Or press 
 and enter the segment number followed by 
.
3. Press 
 and then press either the 
 or 
 key, 
depending on which part of the segment you want to edit.
Cal
 PWRMTR CAL   LOSS/SENSR LISTS   CAL FACTOR SENSOR A 
 ADD     FREQUENCY 
G/n
M/
µ
k/m
 CAL FACTOR 
x1
 DONE 
 DONE 
Cal
 PWRMTR CAL 
 LOSS/SENSR LISTS   CAL FACTOR SENSOR A 
 CAL FACTOR SENSOR B 
 SEGMENT 
 SEGMENT 
x1
 EDIT 
 FREQUENCY 
 CAL FACTOR